Various representations of the resulting 3D spectrum: |
---|
|
Details:
Pulse sequence: standard Bruker pulse sequence hnco3d modified to replace selective shaped pulses on carbon with hard pulses. Spin system: consists of 6 nuclei: Two isolated pairs of 1H-15N-13CO systems, 1H signals at 7 and 9 ppm, 15N at 109 and 111 ppm, and 13C at 159 and 161 ppm, respectively; the scalar couplings were set to JNH = 92 Hz, JNCO = 15 Hz.
Spectrometer frequency: 500.13 MHz (1H) Acquisition: 16 increments in t1 x 32 increments in t2 x 32 complex points in t3, 2 scans per data point Processing details: 2x zero filling, apodization using 90o shifted sinebell, TPPI in both indirect dimensions